Measurement parameters Silicon concentration
Measurement range Silicon concentration range:
0.5-1000 PPb
Resolution 0. 001
Accuracy Silicon concentration:
±1PPb or ±5.0% of Reading,
take a big value of two;
Calibration method Automatic / manual
Zero or slope calibration
Transmitter
Power supply 100-240V AC Frequency50-60Hz
Current output 8 X 0/4-20mA;
Material Antistatic PVC+
heat-resistant plexiglass
Display screen LCD backlight
Language Chinese or English
Sensor
Model SIAse100
Measurement mode Photoelectric colorimetry
Channel distributor
Channel number 1-8
Measurement cycle 10 minutes per channel
Installation requirements
Ambient temperature -20-40℃
Relative humidity 0-80%, No condensation
Protection degree IP65
Pollution level 2
Installation method Wall mounting or panel mounting
Process connection Inlet: 4 X 6 mm PE hose
Outlet: 10 X 12 mm PE hose
Outline dimension 820X480X180mm ( H X W X D )
Weight 18 Kg (body) +8 Kg (reagent bottle)
Sample water requirements
Sample water pressure Inlet: 0-6Bar
Outlet: Atmospheric pressure
Sample water temperature 5-40℃
Sampling velocity 160-320ml/min
Dissolved solid Less than 10PPm without grease